When performing vacuum method leak tests on equipment such as large chambers or long piping, the detector often can't provide sufficient pump evacuation capacity by itself. In this case, the equipment being tested must be evacuated by its own evacuation system or by a separate evacuation system. HELIOT model 103 is a leak tester for equipment maintenance using the vacuum method. It does away with inadequate rough evacuation pumps, and offers high-precision testing in a compact/space-saving, low-cost design.
| Model overview | Maintenance type |
| Sensitivity (Pa·m3/s) | 10-9 to 10-4Pa·m3/s |
| Units | Pa·m3/s, mbar·L/s |
| Main pump/fore pump | Compound molecular pump/membrane pump |
| Sniffer pump | Membrane pump |
| Analyzer tube | Magnetic field deflection mass spectrometer tube |
| Operation | (1) Fully automatic startup (including sensitivity calibration) (*1) (2) Automatic valve operation (3) Calibration leak automatic temperature compensation (4) Quick startup function (5) Zero tracking function (6) Self-check function |
| Leak rate display | Touch-panel display : Digital value readout display (Japanese kanji or English can be selected.) : Analog display (chart/graph display) |
| Leak rate alarm | Audio alarm is provided standard. |
| I/O | RS-232C |
| Printer (Centronics interface) | |
| Recorder (Linear : 0 to 10 VDC with range output; Analog : 0 to 10 VDC) |
|
| External control I/O (device start/stop, test start/end, set points, etc.) | |
| Power supply | AC100V 150W standard |
| Safety standard | CE |
| External dimensions, weight | 260 x 480 x 260mm, 14.5kg |
*1 Sensitivity calibration is performed by internal calibration leak.